EMI Limitations

Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The LOGIQ™ 100 PRO
complies with limits as stated on the EMC label. However there is no guarantee that interference
will not occur in a particular installation.
Possible EMI sources should be identified before the unit is installed.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect.
These sources include:
The presence of a broadcast station or broadcast van may also cause interference.
See Table 2-4 for EMI Prevention tips.
Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 4 2 o f 1 9 7
3
EVISION
medical lasers,
scanners,
cauterizing guns,
computers,
monitors,
fans,
gel warmers,
microwave ovens,
light dimmers
portable phones.