GE LOGIQ 100 PRO Technical Publication page 137

Table of Contents

Advertisement

GE H
EALTHCARE
D
5305880-100, R
IRECTION
8-3-2-2
Optical Trackball Assembly
Procedure for LOGIQ™ 100 PRO
Assemble TB Base Pad2 2222115. Refer Figure 8-5
Note:
Figure 8-5 Optical Trackball Assembly for LOGIQ™ 100 PRO
Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 3 8 o f 1 9 7
3
EVISION
9.) Assemble Optical Trackball PCB Assembly 2224019 refer Figure 8-5
10.)Ensure that Trackball alignment is OK on the Key sheet.
11.)Assemble Trackball bracket assembly 2224020 refer Figure 8-5
12.)Fix screws a,b & c at the location a, b &C in keyboard base plate.
13.)Fix the FPC cable 2139504 (C) as shown in Figure 8-5
14.)For assembly the keyboard PCB follow the reverse order of Section 8-3-2.
When assembling the PCB to the base plate, M3x6 screws only should be used
because a wrong screw used will pierce the key sheet.
15.)For initializing the Optical Trackball Select Alps Trackball from the Trackball Selection
Menu in Service Diagnostics
Chapter 8 - Replacement Procedures
LOGIQ™ 100 PRO S
ERVICE
C
M
ANUAL
Page 8-9