Section 4-3 Functional Check Procedures - GE LOGIQ 100 PRO Technical Publication

Table of Contents

Advertisement

GE H
EALTHCARE
D
5305880-100, R
IRECTION

Section 4-3 Functional Check Procedures

To perform these tests, you will need a convex or a linear probe. These procedures should be per
formed during installation. These procedures are also used as basic checks to use when service of
the system is required.
Turn "ON" the LOGIQ™ 100 PRO system. The 2 LEDS, External Video and Freeze blink and go off.
The Monitor displays the following. Refer figure 4-3 for details.
1.) B-Mode Image appears
2.) Gray Scale Display
3.) Horizontal Scale Marker
4.) Vertical Scale Marker
5.) Hospital Name (If previously entered)
6.) Date
7.) Time
8.) Probe Type
9.) Focus Marker on vertical scale
10.) Depth (150mm)
11.) Gain (G54)
12.) Dynamic Range (D54)
13.) Frequency **.
*
Points 10-12 form a continuous code in the lower left hand corner of the monitor and they are
factory default values.
** Only for C36 Probe.
Page 4-4