Section 10-5 System Periodic Maintenance; Preliminary Checks - GE LOGIQ 100 PRO Technical Publication

Table of Contents

Advertisement

G
HE
E
ALTHCARE
D
5305880-100, R
IRECTION

Section 10-5 System Periodic Maintenance

10-5-1

Preliminary Checks

The preliminary checks take about 15 minutes to perform. Refer to the system user documentation
whenever necessary.
Step
Item
1
Ask & Listen
2
Paperwork
3
Power up
4
Probes
5
Displays
6
Presets
Table 10-6 System Preliminary Checks
Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 7 8 o f 1 9 7
3
EVISION
Ask the customer if they have any problems or questions about the equipment.
Fill in the top of the Periodic Maintenance (PM) Inspection Certificate. Note all probes and system
options.
Turn the system power on and verify that all fans and peripherals turn on. Watch the displays during
power up to verify that no warning or error messages are displayed.
Verify that the system properly recognizes all probes.
Verify proper display on the monitor.
Take printout of all customer presets as a backup.
Chapter 10 - Quality Assurance
LOGIQ™ 100 PRO S
Description
M
ERVICE
ANUAL
Page 10-7