Data sheet for Probe Source Leakage Current
The test passes when all readings measure less than the values shown in Table 10-14 and
Table 10-15. Record all data on the Planned Maintenance Inspection Certificate.
Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit
is powered ON. Be sure to turn the unit power OFF before switching them using the POLARITY
switch and/or the NEUTRAL switch. Otherwise, the unit may be damaged