Section 8-3KEYBOARD

8-3-1

Detaching the Keyboard

Follow the steps below and refer Figure 8-2. For FRU Details refer 9-1-3
1.) Remove Front Panel. Refer 8-4-1.
2.) Remove 14-pin connector from the Davinci.
3.) Remove the black ground cable coming from the keyboard along with the 14 pin flat cable
4.) Remove the TGC connector from the Davinci.
5.) Place a rubber mat below the unit and tilt the unit upward. Remove two screws (a,b) from
6.) Press the lock release and turn the keyboard 90
7.) Remove the 2 ring bushes refer Figure 8-2.
8.) For assembly follow the reverse order.
Service Manual
Table 8-3 Functional tests
Page 8-4
Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 3 3 o f 1 9 7
3
EVISION
connected to the chassis.
the bottom hinge of the keyboard and remove the bottom hinge cover refer Figure 8-2.
carefully pull out the cable with the connector.
Section
Functional Test/ Diagnostic Test
4-2-1
Power ON/ Boot up
4-2-1
Power OFF/ Shut down
4-3
Functional Checks
7-2-3
Keyboard Diagnostics tests
Section 8-3 - KEYBOARD
LOGIQ™ 100 PRO S
ERVICE
°
. Move the keyboard downward and
Debrief Script
Service manual direction 5305880-
100, Section 8-3-1.Equipment
passed all required tests and is
ready for use.