Section 7: Troubleshooting

Replace micro-fuses

Safety gear:
• Disposable lab gloves
Materials:
• Flat blade screwdriver
• Micro-fuses (slow-blow 1.6A)
DANGER!
Risk of death due to electrical current!
Figure 110 Open micro-fuse cover
Figure 111 Remove micro-fuses
Figure 112 Install micro-fuses
1. Power down device using on/off switch on rear side and disconnect power
plug.
2. Open cover (1) using a flat blade screwdriver.
3. Remove micro-fuse attachment.
4. Remove micro-fuses (1) from attachment.
CAUTION!
Risk of property damage!
The unit has two-phase protection. Use only the following fuses:
Slow-blow HBC fuses (1.6A) with a switching capacity of 1500A.
5. Push the micro-fuses included with the delivery into the attachment.
83