Installation, Operation, and Performance Qualification
Terminal Software / LIMS
If CSV format is selected, during start-up or when closing the menu for log format or log port selection, the
device will send the following two lines:
FP6000;Main:VX.XX;COM:VX.X;D:VX.XX;TEC:VX.X<CR>
"batch";"sample";date;value;"dimension";"device-no";"check";"message"<CR>
The first one is an intro line containing version information, the second one is a header line with column
titles for the next result lines.
After each measurement a result line is send which contains the following semicolon-separated values in a
fixed order:
user
batch
sample
date
value
device-no
check
message
Perform some measurements, print your received file and attach the print-out to this page to document the
proper transmission.
COMMENTS
..........................................................................................................................................................................
..........................................................................................................................................................................
Date ............................
Completed by ..............................................
Date ............................
Verified by ..............................................
Column
Description
optional user ID; only present if
user login is activated
Batch identifier in quotation marks or void if batch
ID is disabled in the options (alphanumeric).
Sample identifier in quotation marks or void if
sample ID is disabled in the options (alphanumeric).
Date and time in combined ISO 8601
format (e.g. 2018-12-31T13:45).
Measurement result as integer (in
mOsmol/kg) or error identifier.
Serial number of the device in
quotation marks (alphanumeric).
MD5-Checksum
Human-readable message (always in
English) in quotation marks or void.