Section 8: Servicing

8.6 Perform visual inspection of device

Personnel: User
Safety gear:
• Disposable lab gloves
Materials:
• Fiberglass pin
• Soft cloth
• Tweezers
Figure 139 Inspect device for foreign objects
Figure 140 Clean cooling pin
1. Verify that the cables at the rear side of the device are secure and none of
them are broken.
In case of cable breaks, replace the cable.
2. Check if foreign objects, such as measurement vessels, are located in the
openings of the device.
3. Clean cooling pin (1) using the fiberglass pin. For this purpose, move
elevator cover up.
4. Wipe off dirt and dust from housing using a soft cloth.
5. Move elevator cover down.