Section 2: Design and Function
Error messages
The following error messages may be used to diagnose an operating error or device failure:
Notification
ABORT
User cancels by pressing the EXIT key.
LIFT
User cancels by lifting the sensor.
CANTCOOL
Failed to sufficiently cool down sample.
NOCRYST
Crystallization failed.
SPONCRYST
Spontaneous crystallization of sample.
Checksums
The checksum for each result line is calculated from the contents of the values from the columns Batch ID,
Sample ID, Date/time, Measurement, Unit and Device number. A possible result line:
<Measurement>
<SampleId>SAMPLE01</SampleId>
<DateTime>YYYY-MM-DDTHH:MM:SS</DateTime>
<Value>299</Value>
<Unit>mOsmol/kg</Unit>
<DeviceCode>300201103</DeviceCode>
<CheckSum>756de7e29897809f575ff41fec1a2435</CheckSum>
</Measurement>
...or...
;"SAMPLE01";YYYY-MM-DDTHH:MM:SS;299;"mOsmol/kg";"300201103";
"172ef346c5f36c964ac0710a8421efc1";
...and the above contents are strung together to form:
SAMPLE01YYYY-MM-DDTHH:MM:SS299mOsmol/kg300201103
The MD5 checksum for this string is:
75 6d e7 e2 98 97 80 9f 57 5f f4 1f ec 1a 24 35
See http://en.toolpage.org/tool/md5 to learn more about MD5 calculation.
Meaning
35