Section 7: Troubleshooting
Figure 89 Remove mounting screws
Figure 90 Remove thermistor probe
Figure 91 Remove mounting plate
Figure 92 New thermistor probe with data sheet
8. Move elevator up and remove the 2 mounting screws (1) using Allen key
(SW 2).
9. Remove thermistor probe with mounting plate toward the front while holding
guide tube of initiation needle (1) in place.
10. If necessary, clean cooling pin.
11. Use Allen key (SW 2) (2) to remove thermistor probe from mounting plate
(1).
12. Fit new thermistor probe to mounting plate.
Keep the data sheet affixed to the thermistor probe (2) in the vicinity of the
device.
To protect thermistor, cover thermistor probe with a measurement vessel
(1). The cover of the measurement vessel must point forward.
CAUTION!
Risk of thermistor damage due to loose position of measurement
vessel!
Resistance is felt when sliding on measurement vessel.
When the measurement vessel clicks into place, it is securely
positioned on the thermistor probe.
77