Installation, Operation, and Performance Qualification
Ancillary Information
The correct pipette
Repeatability depends on the performance of the pipette, the used tip and execution of pipetting.
Pipette type: .................................................
S/N:
Calibrated:
Is the pipette equipped with
a disposable tip?
Has been verified that the pipette delivers a
constant sample volume of 15µl or 50µl?*
Is the personnel trained in using the pipette?
*Volume depends on FreezePoint
• Model 6000 / Model 6000P - 15μL
• Model 6000S / Model 6000SP - 50μL
Traceability Osmolality Check
Prior to the traceable osmolality check make sure that the thermistor probe has been properly adjusted
with the adjustment tool according to the user / service guide – especially after changing thermistor probe
or lower cooling system.
The "traceable-check-frequency" has to be carried out in accordance with quality assurance
on site.
COMMENTS
..........................................................................................................................................................................
..........................................................................................................................................................................
Date ............................
Completed by ..............................................
Date ............................
Verified by ..............................................
Comment
.....................................................
.....................................................
.....................................................
.....................................................
.....................................................
.....................................................
Model
®
Date Verified
.....................................................
.....................................................