Section 5: Preventative Maintenance and Safety

5.3 Replacing Fuses

WARNING!
To prevent the risk of fire, the main fuses must only be replaced with fuses of the same type
and rating. Recurring fuse failure indicates serious internal problems, if this occurs, contact
ELITechGroup.
1. Power OFF the instrument.
2. Disconnect the power cord from the power outlet and the rear panel of the instrument.
3. Open the fuse cover by inserting a screwdriver in the slot on the right side of the cover and gently
prying the cover out.
4. Remove the fuse holders to inspect the fuses.
5. Replace the fuses if necessary.
6. Push the fuse holders in.
7. Close the fuse cover.
8. Reconnect the main power cord to the rear panel of the instrument and to the power outlet.
9. Power ON the instrument.