Section 1: Safety

Repeatability of the measurement

Incorrect measurement vessels
NOTE!
Impaired repeatability of measurement due to incorrect measurement vessels!
Repeated use of the measurement vessels and use of incorrect consumables cannot guarantee
reproducible measurement results.
• Always use a clean and unused measurement vessel for every measurement.
• Only use measurement vessels supplied by ELITechGroup.
• Never use centrifuge tubes or reaction vessels.
Improper handling of the calibration standard
NOTE!
Impaired repeatability of measurement due to improper handling of calibration standards!
Improper handling and storage of the calibration standards included with the delivery negatively affects the
measurement accuracy of the device.
• Always observe the stability of the calibration standards after opening the ampule (max. 30 minutes at
72°F (22°C) ambient temperature).
• Never use opened ampules twice or mix them together.
• Never freeze opened ampules.
• Do not use the calibration standards past their expiration date.
Impact/vibration
NOTE!
Increased risk of incorrect measurements!
The device includes precision-engineered components which can be decalibrated and/or damaged
in case of exposure of the device to strong impact/vibration. This can cause a higher risk of incorrect
measurements (spontaneous crystallization).
• Always install the device on a non-vibrating surface.
14