Section 7: Troubleshooting
Figure 102 Move elevator cover up
Figure 103 Remove mounting screws
Figure 104 Remove elevator cover
Figure 105 Install adjustment tool
Figure 106 Align the thermistor probe
4. Move elevator cover up.
5. Remove the 2 mounting screws (1) using Allen key (SW 2).
6. Remove the complete elevator cover.
7. Move the elevator up using handle.
8. Loosen the 2 mounting screws only slightly.
The thermistor probe must be aligned using the adjustment tool.
9. Install adjustment tool (1) in opening of lower cooling system using
tweezers.
10. Remove measurement vessel from thermistor probe.
11. Align thermistor probe. Proceed as follows:
just beyond
Push thermistor probe
12. Position thermistor probe on XY plane so that thermistor is perpendicular to
center of hole of adjustment tool ("Figure 107 Thermistor probe (A) Z axis
and (B) XY plane").
13. Tighten the 2 mounting screws on the aligned thermistor probe.
adjustment tool.
81