Section 10: Specifications
10.1 Dimensions
Specification
Weight
Width
Depth
Height
10.2 Performance parameters
Measuring range
Resolution
Temperature Effects
Precision (within run repeatability)
Models 6000 & 6000P
Sample Volume
0 to ≤ 400 mOsmol/kg
> 400 to ≤ 1500 mOsmol/kg
> 1500 to ≤ 3000 mOsmol/kg
Models 6000S & 6000SP
0 to ≤ 400 mOsmol/kg
> 400 to ≤ 1500 mOsmol/kg
> 1500 to ≤ 3000 mOsmol/kg
SD: standard deviation, CV: coefficient of variation (= relative SD)
Instrument Specifications
0 to approx. 3000 mOsmol/kg H
1 mOsmol/kg over the entire measuring range
Less than 1 mOsmol/kg per 5°C (9°F) ambient temperature change
Range
Value
14.3 (6.5)
8.1 (205)
8.7 (220)
14.2 (360)
O
2
SD ≤ 4 mOsmol/kg
CV ≤ 1%
CV ≤ 2%
SD ≤ 2 mOsmol/kg
CV ≤ 0.5%
CV ≤ 1%
Unit
lbs (kg)
inch (mm)
inch (mm)
inch (mm)
Repeatability
Repeatability
99