Installation, Operation, and Performance Qualification
Performance Specifications
Analytic sensitivity and specificity
Measuring range
Resolution
Temperature Effects
Precision (within run repeatability)*
The osmometer works internally with a nearly tenth as high measuring precision related to the measuring
value. The measuring value is rounded up from 0.5 onwards. This leads to a total deviation from the
expected value. The indication is performed in percentage.
Models 6000 & 6000P
0 to ≤ 400 mOsmol/kg
> 400 to ≤ 1500 mOsmol/kg
> 1500 to ≤ 3000 mOsmol/kg
Models 6000S & 6000SP
0 to ≤ 400 mOsmol/kg
> 400 to ≤ 1500 mOsmol/kg
> 1500 to ≤ 3000 mOsmol/kg
SD: standard deviation, CV: coefficient of variation (= relative SD)
Sample Volume
Range
0 to approx. 3000 mOsmol/kg H
1 mOsmol/kg over the entire measuring range
Less than 1 mOsmol/kg per 5°C (9°F)
ambient temperature change
Repeatability
SD ≤ 4 mOsmol/kg
CV ≤ 1%
CV ≤ 2%
Repeatability
SD ≤ 2 mOsmol/kg
CV ≤ 0.5%
CV ≤ 1%
Value
O
2