Section 6: Operation
Figure 67 Calibration results
Figure 68 Calibration Results
Figure 69 Replace probe
21. Verify the calibration results.
If necessary, tap the individual measurement values to ignore them during
mean value calculation.
(
"10.2 Performance parameters" on page 100)
22. Tap Apply to complete the calibration.
The device is now calibrated.
23. Verify repeatability (
98)
or
Continue with your measurements (
on page 54).
Note regarding 2-point calibration:
For a 2-point calibration, the measurement accuracy of the device
depends on the correct internal calibration of the thermistor.
If the calibration check reveals above average deviations from the
linearity, the lab supervisor has to recalibrate the internal thermistor:
For this purpose, open the Replace Probe menu (Start menu
Options
Maintenance
Tap Calibrate and perform a 3-point calibration using distilled water
and the two calibration standards with 300 and 850 mOsm/kg.
"Check repeatability of measurements" on page
"Measure individual/batch samples"
Replace Probe).
Lab
63