Elitech FreezePoint 6000 Series User Manual page 127

Freezing point osmometer
Table of Contents

Advertisement

Installation, Operation, and Performance Qualification
Accuracy (Linearity)*
3-point calibration
Models 6000 & 6000P (15µl)
mOsmol/kg
0 to ≤ 400
> 400 to ≤ 1000
> 1000 to ≤ 2000
Models 6000S & 6000SP (50µl)
mOsmol/kg
0 to ≤ 400
> 400 to ≤ 1000
> 1000 to ≤ 2000
*Performance above 2000 mOsmol/kg dependent on selection of calibration values. For best results, it is recommended to calibrate
using values near the intended measuring range. Accuracy (from nominal value, tolerances of reference and calibration solutions
excluded) and precision (within run) specifications apply to ELITechGroup standards and reference solutions.
**Measuring points in this range not recommended for indicated calibration values.
COMMENTS
..........................................................................................................................................................................
..........................................................................................................................................................................
Date ............................
Completed by ..............................................
Date ............................
Verified by ..............................................
Calibrated Range
≤ ±4 mOsmol/kg
≤ ±1.0%
**
Calibrated Range
≤ ±2 mOsmol/kg
≤ ±0.5%
**
0 - 1000 mOsmol/kg
Cal. (0/300/850)
0 - 1000 mOsmol/kg
Cal. (0/300/850)
0 - 2000 mOsmol/kg
Cal. (0/850/2000)
≤ ±10 mOsmol/kg
≤ ±3.0%
≤ ±1.5%
0 - 2000 mOsmol/kg
Cal. (0/850/2000)
≤ ±6 mOsmol/kg
≤ ±1.5%
≤ ±1.0%
127