Section 3: Delivery, packaging, and storage

3.2 Unpacking and Device Storage

Unpacking device

1. Carefully open cardboard box using a knife.
2. Remove device from packaging and place in a suitable location.
3. Keep packaging for possible later return.
4. After device has been shipped, please align the thermistor probe upon receiving, as it can become
misaligned in transit and could cause inaccurate readings (
80)

Device Storage

Store the device under the following conditions:
• Do not store outside.
• Store under dry and dust free conditions.
• Do not expose to aggressive media.
• Protect from sunlight.
• Protect the thermistor probe using a measurement vessel.
• Avoid mechanical shock.
• Storage temperature: -10°C to 60°C.
• Relative humidity (non-condensing): 80%.
• If stored for more than 3 months, regularly inspect all parts and packaging for general integrity.
"Align the Thermistor probe" on page
39