Replace Printer Ribbon (Models 6000P / 6000Sp Only) - Elitech FreezePoint 6000 Series User Manual

Freezing point osmometer
Table of Contents

Advertisement

Section 7: Troubleshooting

Replace printer ribbon (Models 6000P / 6000SP only)

Personnel: User
Safety gear:
• Disposable lab gloves
Materials:
• 1 Paper roll
Figure 120 Printer (pulled out)
Figure 121 Pull out printer
Figure 122 Remove printer cover
Figure 123 Push out ribbon
1. Power down device on rear side using on/off switch.
1-Paper roll
2-Ribbon
2. Pull out printer.
3. Remove printer cover.
4. Push on Push marking on front side of ribbon.
The ribbon is released.
5. Remove the released ribbon.