Appendix A: Shipping the device

Personnel: User
Safety gear:
• Disposable lab gloves
When shipping the device for repairs or a refund, please note the following.
Figure 142 Position measurement vessel
Figure 143 Maintenance menu
1. Power down device using on/off switch on rear side and disconnect power
plug.
2. Clean and disinfect the equipment before returning it to us.
Warning
Shipping the device without disinfecting according to instructions is
dangerous to service personnel. You will be charged a additional fees
for cleaning and disinfecting contaminated equipment.
Equipment that is strongly contaminated will not be processed by us
and will be returned at the customer's expense.
3. Position a measurement vessel (1) on the thermistor probe.
NOTE!
The measurement vessel is securely attached to the thermistor probe
when it clicks into place.
4. Move needle to transport position.
Open the menu for configuring the lab options.
In the Start menu, tap Settings
Transport.
The menu may be protected with a lab supervisor pin.
Lab Options
Maintenance
Safe
103