Section 7: Troubleshooting

7.4 Resolve Errors

Replace Initiation needle

Personnel: Service technician
Safety gear:
• Disposable lab gloves
Materials:
• Allen key (SW 2.5)
• Precision key file
• Wire cutter
WARNING!
Risk of infection from sample residue!
The initiation needle is immersed into the sample during measurements. Contact with the initiation needle
increases the risk of infection.
• Wear lab gloves during any kind of work.
DANGER!
Risk of death due to electrical current!
Figure 74 Position of initiation needle and details (#1: Initiation needle)
Figure 75 Remove mounting screw
1. Power down device using on/off switch on rear side and disconnect power
plug.
2. Remove device front panel (
3. Remove mounting screw (1) of initiation needle while holding initiation
needle in place.
NOTE!
The mounting screw has a coating of locking paint.
page 92).