Section 8: Servicing

8.3 Remove device front panel

Remove device front panel
Personnel: Service technician
Safety gear:
• Disposable lab gloves
Materials:
• Allen key SW 2
Figure 126 Position measurement vessel
Figure 127 Move elevator down
1. Power down device on rear side using on/off switch.
2. To protect thermistor, cover thermistor probe with a measurement vessel.
The cover of the measurement vessel must point forward.
CAUTION!
Risk of thermistor damage due to loose position of measurement
vessel!
Resistance is felt when sliding on measurement vessel.
When the measurement vessel clicks into place, it is securely
positioned on the thermistor probe.
3. Move elevator down.