Section 7: Troubleshooting
Figure 81 Check free movement of initiation needle
Figure 82 Move needle
15. Turn motor disk with attached initiation needle clockwise to check initiation
needle for free movement.
If needle gets jammed, detach needle and check if it is bent. (
72).
16. Install device front panel (
17. Power up device on rear side using on/off switch.
18. Verify free movement of initiation needle automatically.
Open the menu for adjusting the initiation needle (Start menu
Options
Maintenance
CAUTION!
Risk of injury from initiation needle!
The tip of the initiation needle is exposed. Movement of the initiation needle
can cause needle puncture injuries.
Always keep your hands and fingers clear from the area underneath the
initiation needle.
19. Tap Move Needle to test the free movement.
The needle turns by one revolution.
Tap Back to exit the menu.
page 95).
Adjust Needle).
page
Lab
74