Elitech FreezePoint 6000 Series User Manual page 59

Freezing point osmometer
Table of Contents

Advertisement

Section 6: Operation
Figure 53 Clean thermistor probe
Figure 54 Perform zero point calibration
Figure 55 Pipette distilled water
Figure 56 Position measurement vessel
1. Clean thermistor probe using a soft, dry, lint free paper towel.
2. Follow the instructions on the touchscreen.
3. Pipette the appropriate volume of distilled water into an unused and clean
measurement vessel.
• Model 6000 / Model 6000P - 15μL
• Model 6000S / Model 6000SP - 50μL
NOTE!
Risk of incorrect measurement resulting from the presence of air
bubbles in the sample!
The sample must be pipetted without air bubbles.
4. Position measurement vessel on thermistor probe with cover facing front.
NOTE!
The measurement vessel (1) is securely attached to the thermistor
probe when it clicks into place.