SYSTem Subsystem

:SYSTem:DATE <year>,<month>,<day>
:SYSTem:DATE?
:SYSTem:ERRor?
Attribute Summary
Description
Query Response
Example
The commands in this subsystem have the following
command hierarchy:
:SYSTem
:DATE
:ERRor?
:HELP
:HEADers?
:PRESet
:TIME
:VERSion?
Sets the clock date. An illegal date will generate an
error and will not set the date.
The query returns the year, month and day.
Queries an error from the error queue.
Preset State: none
*RST State: none
SCPI Compliance: standard
Query Only
The Agilent 86121A has a 30 entry error queue. The
queue is a first-in, first-out buffer. Repeatedly sending
the query :SYSTEM:ERROR? returns the error numbers
and descriptions in the order in which they occur until
the queue is empty. Any further queries returns +0,
"No errors" until another error occurs.
For a complete list of error messages,
Messages" on page
<value>, <string>
<value> is an integer. <string> is the text of the error
message. The following is an example of a response:
-113,"Undefined header"
DIM Error$[250]
OUTPUT 720;":SYSTEM:ERROR?"
8-3.
SYSTem Subsystem 5-57
refer to "Error