Error Messages

Table 8-2. Instrument Specific Error Messages (1 of 3)
Error
Error Message
Number
1
"Bad checksum from motor"
2
"Motor communication problem"
3
"Motor not moving"
4
"Motor index pulse not found"
5
"Motor chip set init failed"
6
"Motor commutation failure"
7
"Motor not settled"
8
"Motor did not stop"
9
"Motor motion error"
10
"Motor position limit failed"
11
"Motor position wrap failed"
12
"Power level too high"
13
"Data download problem"
14
"Data acquisition problem"
15
"Max number of signals found"
16
"Motor interrupt received"
17
"ROM byte unerased"
18
"ROM write operation failed"
19
"ROM defective"
20
"ROM data invalid"
21
"ROM version incompatible"
22
"ROM polling limited out"
23
"Input out of range"
24
"Bad cal ROM data"
25
"Bad cal ROM data"
In this section, you'll find all the error messages that the
Agilent 86121A can display on its screen.
page 8-3
lists all instrument-specific errors.
on page 8-6
lists general SCPI errors.
Description
bad checksum from motor chipset
motor Ready signal timeout
motor did not change position
index pulse not detected
failed chip set initialization
failed motor commutation init
motor timed out before settling
motor did not stop as expected
motor position error exceeded lim
position limit interrupt occurred
position wrap took too long
overrange on the ADC
FIFO download timed out
filling FIFO timed out
found max number of peaks
motor interrupt received
programming an unerased byte
read byte != written byte
ROM may be defective
ROM data is invalid
ROM version incompatible with FW
ROM polling limit reached
input parameter out of range
bad mtr ctl filter const in ROM
bad mtr ctl profile const in ROM
Error Messages
8-3
Table 8-2 on
Table 8-3

Advertisement

Table of Contents