Agilent Technologies 86121A User Manual page 142

Wdm channel analyzer
Table of Contents

Advertisement

Attribute Summary
Description
:CALCulate2:WLIMit:STOP[:WAVelength] {?|{<real> | MINimum | MAXimum }}
Attribute Summary
Constant
MINimum
MAXimum
Non-sequential command
Preset State: 236.0571 THz
*RST State: 236.0571 THz
SCPI Compliance: instrument specific
This command sets the stopping range for the wave-
length limit. The default units for the <real> parameter
are Hz. The stop frequency value must be greater than
or equal to the start frequency value or the stop fre-
quency will be clipped to the start frequency and a
"Data out of range" error will be generated.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that
this command has the time to complete before sending any
more commands to the instrument.
Agilent 86121A to wait for non-sequential commands" on
page 4-11
for more information.
Sets the stopping wavelength for the wavelength limit
range.
<real> is a wavelength value that is within the following
limits:
Constant
MINimum
MAXimum
Non-sequential command
Preset State: 1650 nm
*RST State: 1650 nm
SCPI Compliance: instrument specific
CALCulate2 Subsystem 5-27
Description
start wavelength limit
236.0571 THz (1270 nm)
Refer to "Always force the
Description
start wavelength limit
1650.0 nm