Printing Measurement Results

The Agilent 86121A includes an internal thermal
printer. Measurement results can be sent directly to the
internal printer or to an external printer.
To create a hardcopy on the internal printer
To print using the internal printer, make sure that the
printer contains paper. For information on how to load
paper, see
"Changing the Printer Paper" on page
for information on cleaning the printer head,
"Printer Head Cleaning Procedure" on page
The printer output is not a copy of the display. Rather, it
is a listing of all signals present at the input (up to 200).
The measurement values printed depend on the set-
tings of the instrument when the Print key is pressed.
The following is an example of a typical printout:
17-Sep-1999/14:21:11
Agilent 86121A SER US39400020
Firmware Ver. 1.000
List By Wavelength
8 Lines
Power Offset 0.0 dB
Vacuum
Elevation 0 Meters
Update Normal
Peak Excursion 15 dB
Peak Threshold 10 dB
Device Narrow
Input
Wavelength
Power
1280.384nm
-16.97dBm
1281.473
-13.14
1282.569
-13.92
1283.651
-13.34
1284.752
-11.69
1285.840
-8.11
1286.944
-10.38
1288.034
-14.65
Printing Measurement Results 3-29