5-14 Measurement Instructions
<expected_value>
Constants
<resolution>
Constants
Examples
Query Response
CONFigure command
When this function is used with the CONFigure command, the
query question mark character "?" must not be included in the
string. However, the FETCh, READ, and MEASure command are
queries and require the question mark. Refer to the examples
for this command.
MAXimum
MINimum
DEFault
MAXimum
MINimum
DEFault
:CONF:ARR:POW:FREQ DEF, MIN
:FETC:ARR:POW:FREQ? DEF, MAX
:READ:ARR:POW:FREQ?
:MEAS:ARR:POW:FREQ?
:CONF:SCAL:POW:FREQ 230.8THZ, MAX
:FETC:SCAL:POW:FREQ? 230.8THZ, MIN
:READ:SCAL:POW:FREQ? 230.8THZ
:MEAS:SCAL:POW:FREQ? 230.8THZ
The following line is an example of a returned string
when :MEAS:SCAL:POW:FREQ? MAX is sent:
+1.94055176E+014
If six laser lines are located and
:MEAS:ARR:POW:FREQ? is sent, the following string is
an example of the returned data. Notice that the first
returned number indicates the number of laser-line val-
ues returned in the query.
6,+1.94055100E+014,+1.93854100E+014,+1.93653000E+014,+1.93452000
E+014,+1.93250900E+014,+1.93050000E+014
The highest frequency signal
The lowest frequency signal
The current marker position
0.01 resolution (fast update)
0.001 resolution (normal)
Current resolution