Agilent Technologies 86121A User Manual page 170

Wdm channel analyzer
Table of Contents

Advertisement

:STATus:OPERation:NTRansition{?| <integer>}
Attribute Summary
Description
Example
:STATus:OPERation:PTRansition{?| <integer>}
Attribute Summary
Description
Selects bits in the event register which can be set by
negative transitions of the corresponding bits in the
condition register.
<integer> an integer from 0 to 65535.
Preset State: none
*RST State: none
SCPI Compliance: standard
Changes in the state of a condition register bit causes
the associated OPERation Status or QUEStionable Sta-
tus register bit to be set. This command allows you to
select a negative bit transition to trigger an event to be
recognized. A negative transition is defined to occur
whenever the selected bit changes states from a 1 to a
0. You can enter any value from 0 to 65535.
When queried, the largest value that can be returned is
32767. This is because the most-significant register bit
cannot be set true.
OUTPUT 720;":STATUS:OPER:NTRansition 16"
Selects bits in the event register which can be set by
positive transitions of the corresponding bits in the con-
dition register.
<integer> an integer from 0 to 65535.
Preset State: none
*RST State: none
SCPI Compliance: standard
Changes in the state of a condition register bit causes
the associated OPERation Status or QUEStionable Sta-
tus event register bit to be set. This command allows
you to select a positive bit transition to trigger an event
to be recognized. A positive transition is defined to
occur whenever the selected bit changes states from a
0 to a 1. You can enter any value from 0 to 65535.
STATus Subsystem 5-55

Advertisement

Table of Contents
loading