:CALCulate2:POINts?
Attribute Summary
Description
Query Response
:CALCulate2:PTHReshold{?| {<integer> | MINimum | MAXimum | DEFault}}
The query response is the current value. For example, if
the current value is set to 15 dB, the following value is
returned:
15
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that
this command has the time to complete before sending any
more commands to the instrument.
Agilent 86121A to wait for non-sequential commands" on
page 4-11
for more information.
Queries the number of points in the data set.
Preset State: unaffected
*RST State: unaffected
SCPI Compliance: instrument specific
Query Only
This is the number of points that will be returned by the
CALC2:DATA? query.
For example, if six laser lines are located:
+6
Sets the peak threshold limit used by the instrument to
determine valid laser line peaks.
<integer> represents logarithmic units in dB. Valid
range is 0 to 40.
Constant
MINimum
MAXimum
DEFault
CALCulate2 Subsystem 5-23
Refer to "Always force the
Value
0 dB
40 dB
10 dB