Preparation

Doors, ports and bed

Doors, ports and bed
Access ports
To open the access port:
1
Press down the knurled area of the locking bolt. The access
port opens.
To close the access port:
Press the access port closed until the locking bolt engages.
Front flap
To open the front flap:
2
Turn the two knobs inwards to the vertical position.
The red latch becomes visible.
Lower the front flap until it hangs down vertically towards
the floor.
Take care not to trap hoses and cables in the moveable double
wall of the front flap!
To close the front flap:
Raise the front flap and press into position,
3
Turn the two locking knobs outwards to the horizontal
position until you feel them click into place.
WARNING
Flap can open on its own, danger to patients.
Make sure that both locking knobs are engaged. The red
latches must no longer be visible.
WARNING
Danger of excessive cooling or overheating of the
patient. If the moveable double wall is not parallel to the
front flap, the hot air duct is interrupted.
The moveable double wall must be positioned parallel
to the front flap. Align the moveable double wall if neces-
sary!