Operation
Precautions
WARNING
Hoses and/or cables risk being trapped when tilting the
®
Caleo
and when opening and closing the front flap.
Ensure that all hoses and cables are routed correctly and
safely without obstruction!
NOTE
Risk of damage to control unit due to improper handling
Position the control unit so that the screen is clearly visible
from the working area. Do not mechanically load the cable by
e.g. hanging objects from it. If the control unit is mounted on
the handle side, never remove the cable to the control unit
from the cable guides on the basic pole.
Do not place any liquid containers above the control unit.
CAUTION
Risk of damage to Caleo or accessories when adjusting the
tilt angle and height.
When fitting accessories, please note: The accessory equip-
ment must not collide with the unit when adjusting the tilt
®
angle and height of Caleo
NOTE
Risk of damage to Caleo or accessories when adjusting the tilt
angle and height.
When fitting accessories, please note: The accessory equip-
ment on the handle side does not move with the main unit
during height adjustment of Caleo
accessories such that there is sufficient free space.
WARNING
Risk of tilting over if the maximum load of the accesso-
ries is exceeded.
Observe the maximum loads of the accessories.
WARNING
Kangaroo mode
Danger to patient of hypothermia or overheating.
The patient's core temperature must be monitored
constantly as it is not part of the automatic temperature
control.
Particular attention must be paid to vital parameters of
critical patients.
42
.
®
(see page 123). Position
When using Phototherapy
WARNING
Danger to patient of overheating.
Absorption of light through the patient's skin will supply
heat. This may increase the core temperature.
Monitor the core temperature of the patient regularly with
particular care during phototherapy.
WARNING
Reduce the set value of air temperature approx.
15 minutes before starting phototherapy by approx. 2
Decrease the set value for humidity.
The room temperature must be at least 3
the air temperature in the Caleo
applies to Dräger Type 4000 phototherapy units.
When using other phototherapy units, especially units
without a built-in fan, the temperature in the Caleo
bator may rise even more sharply.
Phototherapy may only be used with a stand.
WARNING
Fluid balance can be affected adversely by phototherapy.
The supply of fluid to the patient must be increased, e.g.
by parentral infusion, to compensate for the increased
loss of water during phototherapy.
WARNING
The phototherapy light and Caleo
covered with cloths, aluminium foil or other materials in
order to boost the phototherapeutic effect to avoid the
risk of overheating the patient.
Do not cover the phototherapy light and hood of Caleo
with cloth, aluminium foil etc.
Weighing scale
WARNING
If you fail to observe the Instructions for Use (see
page 89), considerable measuring errors may occur
when determining the patient weight.
Only use the built-in scale to determine the weight of the
patient.
To back up critical therapeutic decisions that are based
on the weight of the patient, check the measured weight
with the help of a reference measurement carried out on
an external scale.
Transport within hospitals
WARNING
During internal transport outside the hospital building,
the castors can be damaged or dislodged. Risk of
toppling over.
An even floor surface must be ensured when moving the
Caleo.
®
Do not use Caleo
outside the hospital building.
®
The Caleo
may only be moved when empty.
o
C.
o
C lower than
®
incubator. This value
®
incu-
®
hood must not be