Preparation
Doors, ports and bed
Removing the double wall
1
Pull the plunger up as far as it will go.
The red shaft of the plunger will then be visible.
2
Grip the double wall with both hands and lift it off.
Storing the double wall
Fix the hanger (2M 51 152) to the compact rail of the Caleo.
3
Hook the double wall by its plunger to the hanger.
CAUTION
If Caleo is lowered too far, the double wall may touch the
floor. Risk of damage.
Observe the set height of the Caleo.
Make sure that if the Caleo is lowered the double wall does
not touch the floor.
Canopy
To open the canopy:
4
Grasp the handle of the canopy
and
o
5
open it (approx. 60
).
WARNING
Danger of overturning the canopy.
Do not tilt the canopy over the side. Tip the canopy only
longitudinally, as shown.
6
Raise the side canopy prop, and lower the canopy until the
prop is fixed in the slot in the canopy.
To close the canopy:
4
Grasp the handle on the canopy and raise it slightly.
6
Fold down the prop
and
close the canopy.
The canopy can be opened from both sides.