Cleaning, disinfecting and sterilizing
Stripping down
®
Set the bed of Caleo
to a horizontal position (see page 27).
Switch on cleaning mode
NOTE
*
After Cleaning Mode
has ended or if humidity control is not
available:
Switch off the incubator and disconnect it from the mains
power supply.
Remove any auxiliary equipment installed
(for reprocessing instructions see the specific
Instructions for Use of the equipment concerned).
WARNING
Risk of burning when touching the heater!
When the incubator is closed, the heater is still hot
enough to cause serious burns for a long time after
switching off (still approx. 70
®
Allow Caleo
to cool down sufficiently before stripping
it down.
Removing the double wall
1
Pull the plunger up as far as it will go.
The red shaft of the plunger will then be visible.
2
Grip the double wall with both hands and lift it off.
Place the double wall on a soft, non-scratch surface.
Remove the continuous seal from around the double wall.
3
Force the bore sleeve apart
and
4
remove the plunger.
*
Only available in combination with humidity control.
98
*
(see page 79).
o
C after 1 hour).