Operation
Using skin temperature control

Alarms

Alarm limits can be changed in the configuration (page 86).
If the deviation between the set and measured skin tempera-
o
*
ture exceeds ±0.5
:
C
The alarm message appears on the screen.
»Skin temperature deviation. 1 more than 0.5
The alarm tone sequence (3 beeps) is sounded,
1
The central alarm indicator lights up
2
The measured value flashes,
3
Yellow bar LED flashes.
The alarm tone can be muted for 5 minutes:
4
Suppress alarm tone = press key.
or
5
Press the rotary knob.
The warning message remains on the screen,
The intermittent alarm tone is muted,
1
The central alarm indicator goes out,
2
The measured value continues to flash,
3
Yellow bar LED continues to flash.
If the measured value is in the range ±0.5
The warning message disappears,
The intermittent alarm tone is muted,
1
The central alarm indicator goes out,
2
The measured value remains on-screen but does not flash,
3
Yellow bar LED goes out. If the extended set value range is
used the yellow bar LED will light up.
If the sensor plug is disconnected:
After 3 seconds:
The alarm message appears on the screen.
»Connect skin temperature sensor 1«.
The alarm tone sequence (5 beeps) is sounded,
1
The central alarm indicator lights up**,
2
3 flashing dashes are displayed in place of a temperature
reading,
6
Red bar LED flashes.
Then:
Immediately plug in the sensor.
*
The numerical values in this description are examples. See "Setting
alarms" on page 85.
**
The central alarm indicator can be switched off.
See "Setting system parameters" on page 84.
56
o
C«,
**
,
o
C again:
1
6
3
4
2
5