Dräger Caleo Instructions For Use Manual page 57

Neonatal incubator
Hide thumbs Also See for Caleo:
Table of Contents

Advertisement

WARNING
If the skin temperature sensor is not inserted or if it is
defective, Caleo does not warm up as long 3 dashes are
displayed on the screen. Danger of hypothermia for the
patient.
Inserting or replacing the skin temperature sensor.
If the sensor is defective:
The alarm message appears on the screen.
»Skin temperature sensor 1 error«.
The intermittent alarm tone sequence (5 beeps) is sounded,
The central alarm indicator lights up
1
3 flashing dashes are displayed in place of a temperature
reading,
2
Red bar LED flashes.
Then:
Change the skin temperature sensor.
The alarm tone can be muted for 5 minutes:
3
Suppress intermittent alarm tone = press key
or
4
Press the rotary knob.
The warning message remains on the screen,
The intermittent alarm tone is muted,
The central alarm indicator goes out,
1
3 dashes continue flashing,
2
Red bar LED continues to flash.
Other alarms, see "Message – Cause – Remedy" on page 112.
See "Alarm descriptions" on page 138.
*
The central alarm indicator can be switched off.
See "Setting system parameters" on page 84.
*
,
Using skin temperature control
2
3
1
4
Operation
57