Changing over between air/skin temperature
control
WARNING
For babies who are in a state of shock, skin temperature
control would increase the air temperature too much and
would endanger the patient, as in this case the skin
temperature is considerably lower than normal.
Do not use skin temperature control on patients who
are in a state of shock. Use air temperature control,
see page 45.
WARNING
For patients who have fever, skin temperature control
would cool the air temperature too much and would
endanger the patient, as in this case the skin tempera-
ture is considerably higher than normal.
Do not use skin temperature control on patients who
have fever. Use air temperature control, see page 45.
WARNING
When treating twins in Caleo, the skin temperature
control monitors the temperature of one patient only.
Danger of hypothermia or overheating.
Do not use skin temperature control, use air temperature
control, see page 45.
1
Change temperature control mode = press key.
1
The LED for the control mode requiring confirmation
flashes.
The activated control mode is displayed on the screen.
The actual and set values are displayed as bar graphs and
numeric values.
The following message appears in the upper section of the
screen.
»set value with rotary knob«

Changing over between air/skin temperature control