Preparation; Before Using For The First Time; Mounting Accessories - Dräger Caleo Instructions For Use Manual

Neonatal incubator
Hide thumbs Also See for Caleo:
Table of Contents

Advertisement

Preparation

Before using for the first time

Mounting accessories

Preparation
Before using for the first time
Check that all packaging materials have been completely
removed (see packing slip in the pack).
Check that the mains power supply voltage matches the
voltage rating specified on the nameplate (see page 140).
Check that the height above sea level is correctly set
(see page 83).
Mounting accessories
CAUTION
When fitting accessories, please note
Any mounted accessories should not collide with the unit
when adjusting the tilt angle and heigh oft Caleo
damage. Install the accessories such that a collision with the
device is ruled out.
Screw on pole 38 mm/600 (2M 50 691) or
NOTE
When fitting accessories, please note:
Accessories mounted on the handle side do not move
with the main unit during the height adjustment of Caleo
(see page 123). Hoses or cables can be stretched during the
height adjustment of Caleo. Lay hoses or cables such that
they are sufficiently long during the height adjustment.
Screw on pole 38 mm/600 (2M50691) or
pole 38 mm/310 (2M 50 688) or
pole 25 mm/600 (2M 50 689).
Remove the cover plate from the base frame.
Screw the pole fully into the base frame and tighten firmly.
Make sure that it is securely held in place.
Observe the maximum loads.
Danger of overturning the unit!
Pole 38 mm/600
10 kg
Pole 38 mm/310
10 kg
Pole 25 mm/600
3 kg
Distance between the loads and the pole: max. 150 mm.
Moving the control unit to the opposite side
for pole, 38 mm
1
Remove the control unit = loosen the clamping screw.
2
Remove the holder = loosen the clamping screw.
On delivery, the control unit is fixed to the 38 mm/310 tube.
This tube is secured by a screwed fixing for transport. If
required, the tube can be released (recommendation: ask for
assistance from specialized technical personnel):
3
Remove both caps from the tube. Insert a screwdriver in the
holes and release the tube.
To fit the control unit to the adjacent side:
Push the holder over the tube and set it to the desired
working height.
2
Mount the bracket = tighten the clamping screw.
Set the control unit to the desired working height.
1
Secure the control unit to the holder = tighten the clamping
screw.
14
®
. Risk of
®
1
2
3

Hide quick links:

Advertisement

Table of Contents
loading