Dräger Caleo Instructions For Use Manual page 77

Neonatal incubator
Hide thumbs Also See for Caleo:
Table of Contents

Advertisement

The newly selected trend 2 is displayed on the screen under
trend 1. In addition, a second data window containing the rele-
vant parameters is opened next to trend 2.
The time cursor and time range (zoom) are identical for both
trend displays.
The »disable« option removes trend 2 from the screen, so that
only trend 1 is displayed.
Selecting the time interval (zoom)
1
Open the Zoom menu = press soft key.
The following intervals of time can be selected for the time
range (zoom) function:
3; 6; 12; 24; 48 hours or 7 days.
2
Select zoom = Turn rotary knob.
2
Activate zoom = Press rotary knob.
The selected trends will then be displayed in the newly
selected time range.
All data going back a maximum of 7 days can be displayed.
The individual measuring points together with the associated
data are displayed as average values in the data window.
When displaying the measured weight values, up to 30 values
can be stored and displayed as a trend. Values between the
individual measured values are interpolated.
Ending trend analysis
To exit trend analysis
Back to menu = press »
or
3
Terminate trend analysis = press key
or
if no soft key on the screen is pressed for 2 minutes,
the display will automatically revert to the standard screen.
« key.
Operation
Selecting menus
3
2
1
77