High noise levels
NOTE
Excessive noise levels that can disturb the patient may be
caused by:
using oxygen hoods and delivering pressurised gases,
wear on the bearings of the fan motor,
Storage of parts on the hood, hence do not place any parts
on the hood.
Observe the specified maintenance intervals, see page 106.
Cleaning Mode
WARNING
Danger to the patient.
Use the Cleaning mode only if Caleo
WARNING
Risk of burning from the heater!
®
Allow Caleo
to cool down before stripping it down.
WARNING
Risk of slipping due to moisture under the Caleo. During
or after the Cleaning mode, condensation could form
under the housing of the Caleo
AUTO humidity can be used to minimise the effect of
condensation (page 135).
®
is empty.
®
!
Electrical safety
WARNING
With add-on devices that do not conform to IEC 60601-1,
the risk of an electric shock cannot be ruled out if there
is an error.
Only use electromedical devices conforming to
IEC 60601-1.
WARNING
Use of socket strips to connect the power supply of the
®
Caleo
may lead to the risk of an electric shock.
The patient leakage current and the ground leakage
current may rise above the permitted limits if the
grounding cable fails when equipment is connected to
the socket strip. The risk of electric shock cannot be
excluded.
Do not use power socket strips to connect the power
®
supply of the Caleo
!
WARNING
When using the integrated socket strip: Non-observation
of the permissible total ground leakage current and total
current consumption can cause the danger of an electric
shock.
Take into account the total leakage current and total
current consumption. See "Technical Data" on page 120.
WARNING
Risk of electric shock from external devices connected
to the interface if Caleo is not connected to the power
supply socket via the power cable or not grounded via
the grounding bolt on the back of the unit.
Only connect external devices to the interfaces if Caleo
is connected to the power supply socket via the power
cable or the device is grounded via grounding bolt on the
back of the unit.
WARNING
Risk to patient or user from external devices that are not
fed by electricity.
Caleo does not monitor the power supply to external
devices.
Monitor the power supply to external devices separately.
Operation
Precautions
43