Preparation
Doors, ports and bed
Trolley with variable height adjustment
To use the height adjustment facility:
®
Switch on Caleo
(see page 44).
1
Press the left pedal – Caleo
2
Press the right foot pedal – Caleo
Adjust to a comfortable working height.
When the height stops changing, the trolley is at its end
position. Release the pedal.
WARNING
Too short hoses or cables can kink, tear or become
squashed.
Hoses and cables must be long enough not to kink, tear
or become squashed!
CAUTION
Objects stored under the drawer can lead to damage.
Do not store anything under the drawer!
NOTE
The height adjustment and bed-tilt mechanism cannot be
operated at the same time! Function interruption.
Activate only the height adjustment or bed-tilt mechanism.
NOTE
Height adjustment is not suitable for steady operation.
Function interruption
The height adjustment can only be operated for max.
6 minutes during one hour.
Trolley castor with direction lock
Activate the lock lever: the castor will be locked in the
lengthways direction (longitudinal axis of device).
The castor with direction lock does not have a brake.
When parking the unit, apply the brakes of all braked
castors.
WARNING
The castor with direction lock does not have a brake.
The device can start moving on its own.
When parking the unit, apply the brakes of all braked
castors.
*
optional equipment feature
26
*
®
is lowered.
®
is raised.
1
2