Communication Ports

1 The Fabius OS has two ports on the back panel
for communication with external devices. The
ports are labeled COM 1 and COM
port MEDIBUS and Vitalink communications.
WARNING
A test for leakage current must be performed
by qualified engineering personnel before use
if the Fabius OS is interfaced with other equip-
ment.
CAUTION
Only the combinations of Dräger approved moni-
tors, mounting parts and interface cables may be
used. Otherwise the correct functioning of the
device may be compromised.

Recommended Device Configuration

NOTE
Data Pass Through (gas analysis data) must be
enabled by DrägerService or your local authorized
service organization.
Configuration 1:
– Fabius OS
– Gas Analyzer
– port COM 2*
2 Connect gas analyzer to either COM 1 or
COM 2*.
*
optional
Instructions for Use Fabius OS SW 3.n
2*
and sup-
1
2
OXYGEN
SENSOR
BREATHING
PRESSURE
VOLUME
SENSOR
System Overview
2
23