Dräger Caleo Instructions For Use Manual page 41

Neonatal incubator
Hide thumbs Also See for Caleo:
Table of Contents

Advertisement

WARNING
Fire hazard from supply of O
Avoid open fire or smoking. Textiles, oil and plastics
can very easily catch fire and burn rapidly in an O
enriched atmosphere.
All fittings and seals in contact with oxygen must be
kept free of oil and grease!
Open valves on O
cylinders slowly.
2
®
Do not use Caleo
in the presence of flammable
anaesthetic gases or disinfecting agents. Risk of
explosion.
Do not use or keep flammable liquids, such as alcohol,
ether or acetone, in Caleo
Do not use any electrical appliances in the patent
capsule unless this equipment is expressly designed
for use in explosion-hazard areas.
WARNING
Physiological risk to patient due to increased O
When making treatment decisions pay attention to physi-
ological risks from O
.
2
NOTE
Physiological risk to patient due to increased O
During O
therapy, monitor the O
2
rate O
monitor.
2
WARNING
Risk to the patient due to of hyperoxemia (possibility of
damage to the eyes) or hypoxemia (possibility of brain
damage) from O
therapy. The air in the incubator should
2
only be enriched with O
2
Oxygen is classified as a drug. O
controlled on the basis of measured O
patient's blood or SaO
or SpO
2
a danger of hyperoxemia (possibility of damage to the
eyes) or hypoxemia (possibility of brain damage).
WARNING
Risk of impairment of device functioning due to the
precipitation of atomized substances.
Medicaments and similar substances must not be
atomized in the patient capsule.
CAUTION
Risk of damage to sensor unit from covering or impurities or
use as fastening point.
Do not cover the sensor unit and do not fasten anything on it.
Keep the slits of the sensor unit free of impurities.
2
®
!
supply.
2
supply.
2
concentration with a sepa-
2
when prescribed by a doctor.
enrichment must be
2
saturation in the
2
). Otherwise, there is
2
WARNING
Physiological risk to patient due to increased CO
concentration.
-
In the event of a power failure, the CO
2
capsule may rise due to an insufficient supply of fresh
air.
Heed the risk of CO
connection.
CAUTION
Alarm cannot be detected via the central alarm indicator if the
alarm is suppressed (see page 139).
Always pay attention to the bar LEDs and the alarm signal on
the control unit.
WARNING
Danger to patient from wrongly actuated front flaps.
When closing the front flaps, make sure that the patient
is not lying in the closing path. The front flaps are not
shut correctly until the red latches are no longer visible.
When opening and closing the front flaps, make sure that
hoses and cables are not caught in the moveable
double wall!
The moveable double wall must be positioned parallel to
the front flap otherwise the warm air duct may be
obstructed!
WARNING
Danger to patient from wrongly actuated side flaps.
When opening and closing the side flaps, make sure that
the hoses and cables are routed safely and clear of any
obstructions!
The side flaps are not shut correctly until the red latches
are no longer visible.
CAUTION
To avoid bacterial contamination when reaching through the
access ports into the patient capsule of Caleo, observe suit-
able hygiene measures and, if possible, do not touch the
surface of the patient capsule.
CAUTION
Danger to patient from wrongly used hood.
The hood must not be used as a shelf for laying clothing,
instruments etc.
Do not hang any objects from the hooks for the double walls.
Before moving the hood, make sure that nothing has been
laid on top of it.
When fitting and removing the canopy, hold it firmly in your
hand. The canopy prop must engage correctly. Do not lift the
hood when the canopy is open. Do not tilt the hood over the
side. When closed, make sure that the canopy sits firmly in
place.
level in the patient
2
poisoning, re-stablish power supply
2
Operation
Precautions
2
41