Cleaning, disinfecting and sterilizing
Stripping down
Base frame
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect
(see manufacturer's specifications), wipe the surface
with a clean damp cloth and then rub dry.
Remove any impurities near the slits of the sensor unit.
CAUTION
When cleaning and wipe-disinfecting the base frame,
penetrating humidity may damage the sensor unit.
Do not allow any moisture to enter the sensor. Do not
disinfect the sensor unit by immersion or spraying.
Control panel
Pull off the rotary knob.
Remove visible soiling with a disposable cloth and
detergent.
Wipe-disinfect the surfaces.
After allowing the disinfectant time to take effect
(see manufacturer's specifications), wipe the surface
with a clean damp cloth and then rub dry.
CAUTION
When cleaning and wipe-disinfecting the control unit,
penetrating moisture may damage the sensor unit.
Do not allow any moisture to enter the sensor. Do not
disinfect the sensor unit by immersion or spraying.
NOTE
Risk of cracks forming in the transparent parts if cleaning
agents and disinfectants not recommended in this Instructions
for Use are used, such as alcohol.
Use only the recommended cleaning agents and disinfectants
(page 105).
NOTE
Crack formation in the plastic components is possible due to
UV light radiation. Do not expose Caleo to UV radiation.