Warming bed
The bed is enclosed by two side panels and two end panels
that can be opened independently of one another.
The side panels are provided with holes for secretion and
drainage hoses.
The end panels have flexible grommets for hoses and cables.
Fitting the panels
1 Insert the lower mounting pins of the panel hinge-pieces
into the guide slots on either side.
With the panels on the long side, ensure that the handle
points outwards.
Position the panel semi-upright,
●
2 press the panel down firmly until the pins lock into position
at the bottom of the slots.
Fold the panel up into the vertical position and allow it to
●
drop into the locking position.
Fit all four panels in this way.
●
Make sure that the panels are seated correctly.
To open:
Lift the panel all the way up until the upper pins come out
●
of their slots, then fold the panel down.
With lively infants or older children:
Use 230 mm high panels.
●
Fitting the inner panels
WARNING
Inner panels should be used with all patients, regard-
less of whether the side panels are up or down!
Otherwise the patient may fall out!
Insert the pins into the tapered holes in the corners of the
●
housing and press down lightly until the inner panel is
firmly in place.
Make sure the panels are seated correctly.
Mattress
Lay gel mattress flat in the cot and cover it with a cotton
●
sheet.
Lay SoftBed in the warming therapy device dimpled side
●
up.
No additional sheet is required for the patient to lie on.
Lay the patient on the dimpled side.
●
2
1
Preparation
2
1
13