Using O 2 - Control - Dräger Caleo Instructions For Use Manual

Neonatal incubator
Hide thumbs Also See for Caleo:
Table of Contents

Advertisement

Operation
Using O2- control
*
Using O
control
-
2
WARNING
Risk to patient due to hyperoxemia (possibility of
damage to the eyes) or hypoxemia (possibility of brain
damage) from O
therapy. The air in the incubator should
2
only be enriched with O
2
Oxygen is classified as a drug. O
controlled on the basis of measured O
patient's blood or SaO
or SpO
2
a danger of hyperoxemia (possibility of damage to the
eyes) or hypoxemia (possibility of brain damage).
Connect the probe of the O
medical gas pipeline supply (page 19).
1
Set O
control = press key.
2
The actual value and the current set value of the O
are displayed as bar graphs and numerical values.
Soft Key assignments:
2
= Do not activate new settings.
3
off
= Switch off O
4
on
= Switch on O
After switching on the device, the oxygen sensors need
a warm-up phase of at least 3 minutes. An hourglass
symbol appears on the screen.
The following message is displayed at the top of the screen:
»set value with rotary knob«.
After confirming the selected set value for O
completion of the warm-up phase, the automatic O
is activated.
The current measured value is displayed after completing the
warm-up phase.
*
optional equipment feature
62
when prescribed by a doctor.
enrichment must be
2
saturation in the
2
). Otherwise, there is
2
hose to the outlet of the central
2
2
control.
2
control.
2
and after
2
2
control
control
1
2 3
4