Preparation
Doors, ports and bed
Mounting accessories
Using humidifier systems
NOTE
Possible malfunction due to use of non-distilled or non-sterile
water.
Only use demineralized water (e.g. Aqua dest.)! Do not use
any additives!
Water tank (2M 50 040):
Disinfect hands.
Open the water tank = lift up the cap.
Fill the water tank with demineralized water (Aqua dest.).
Always fill the water tank to the full level marking.
Capacity: 2.8 L
Do not use any additives!
Close water tank = push down the cap.
Prepare a fresh transfer set (MX 17 018).
Close the clamp on the transfer set.
Pierce the silicone nozzle of the water tank with the pin of
the transfer set.
Open the clamp on the transfer set.
Bleed the transfer set (let sterilised water drain off).
Close the clamp on the transfer set.
Open the clamp on the transfer set.
Switch on the humidification module on Caleo
humidity value (page 59).
Water tank empty = Water shortage alarm is triggered on
®
Caleo
.
Minimum water quality requirements
To avoid impairment of function and long-term material
damages, please use only water with the following minimum
quality for humidication:
Demineralized water (e.g. Aqua dest.) or
Demineralized water with a conductivity <20 µS/cm
Do not use any mineral-containing water, e.g. tap water.
Water heater
The water heater does not have any components that must be
dismantled for the preparation.
Execute the Cleaning Mode after each change of patient.
While preparing the interior of the incubator, ensure that no
liquids such as detergents penetrate into the water heater.
Water bag
NOTE
Use only original sealed bag. Do not use additives or confuse
with infusion solutions. Risk of patient injury.
Only use original sealed bags filled with demineralized water
(e.g. Aqua dest.). Do not use any additives! Do not confuse
with infusion solutions!
Disinfect hands.
Prepare a new transfer set (MX 17 018) and a water bag
with distilled water (e.g. Aqua dest.).
Close the clamp on the transfer set.